In-vitro verification of tolerance limits for biotin interference of routine biochemistry and immunology assays

Swallow K; Wild G; Sargur R

Annals of clinical biochemistry ยท 2026

In plain language

Spiking samples with biotin equivalent to doses of 0.05 to 3600 mg daily showed that not every biotin-based assay is affected, but some were disturbed below manufacturer thresholds and three unwarned assays showed interference.